Q3 2009 / FID Maintenance – Key tips for the in-house detector doctor
Seal the detector inlet in the oven with a detector plug
and ignite the detector. If the chromatogram noise disappears, then the source
of the problem is contaminants in the carrier gas or bleed from the chromatography
column, not a dirty FID.
To prevent this problem, you should always use a carrier
gas purification system and columns prepared from chromatography quality
stationary phases.
Second, check for gas-related noise sources
A problem in the hydrogen or compressed air delivery system can be a source of noise. Measure the flow rates for both gases. As a general rule the air flow should be nearly ten times the hydrogen flow. However, check your instrument’s operators manual for the exact flow requirements of your model. An incorrect flow rate in either gas can cause noise, lack of sensitivity, and/or difficulty when igniting the flame.
Also,
check the connections on both systems. An electronic gas leak detector will
quickly and easily pinpoint gas leaks via differences in thermal conductivity.
If you do not find a major leak, check the gas pressure at the cylinders. If
the pressure is less than 500 psig, replace the cylinder.
A contaminated gas cylinder could be the source of the
problem, especially if the noise appeared several hours after a cylinder
change. Check each cylinder for contaminants and replace if necessary.
Eliminating electrical noise
Electrical noise is another potential problem. It can
cause symptoms similar to a dirty FID. A poor connection due to oxidized
contacts can act as a small capacitor causing spikes and/or loss of
sensitivity. If the FID has clip contacts to the collector or flame jet, clean
them periodically with an emery cloth. Other electrical devices located near
the chromatograph can also interfere with the instrument’s power supply and
cause intermittent or random spikes.
To isolate this source of noise, disconnect the
electrometer cable(s) from the FID. If noise persists, it is coming from the
electrical system. In a two-detector system, electrical noise will appear on
both detectors. If noise is present on only one channel, the electrometer or
electrometer cable(s) are the problem.
Switch the cables at the electrometer and FID. If the
noise now appears on the channel that was previously noise-free, replace the
cable supplying the noise. If the noise does not change channels, check the
electrometer on the malfunctioning channel by removing the electrometer cable
at the FID. Do this by resting the cable connector on a surface to prevent
contact with the metal frame of the GC. If the noise continues, the electrometer
should be replaced. If the noise is eliminated, re-check the hydrogen and air
systems.
Cleaning an FID
Properly cleaning an FID requires cleaning the collector
assembly, the jets, the Teflon insulators, and the housing. If the detector has
been cleaned recently, a simple light coat of silica or even a single silica
flake could cause the trouble, and only a light cleaning is required. If the
detector has not been cleaned recently, a thorough cleaning is necessary.
How to perform a light cleaning
Disassemble the detector and scrub the contamination from
the components. The brass wire brushes will not scratch metal or ceramic parts.
Use a nylon brush on Teflon parts and follow these steps.
1. Disconnect the power to the GC, and be sure the
collector assembly is cool.
2. Remove the collector assembly and brush the collector
to remove the deposits.
3. Replace the jet.
4. Clean the electrical contacts, using a fine emery
cloth. (Be careful not to bend the contacts.)
How to perform a thorough cleaning
If a light cleaning does not succeed, the thorough
cleaning method is required. To accomplish this start by unplugging the power
cord from the electrical outlet, allow the detector to cool, disassemble it,
and follow these steps.
1. Fill the basin in an ultrasonic cleaning device with a
detergent that will effectively remove silica and other contamination from the
FID (e.g., 10:1 water to detergent ratio will clean soiled parts). Immerse the
FID parts, except the electrical contacts, and sonicate for two hours. You can
brush the collector assembly and jets (with nylon or brass brushes) with the
cleaning solution during the ultrasonic treatment. After the treatment, rinse
the parts with distilled water to remove the detergent, and dry parts
thoroughly. Use a fine emery cloth to clean the electrical contacts.
2. After you have cleaned all parts of the detector, check
all O-rings and replace them if necessary. Worn-out O-rings will cause gas
leaks, which can produce detector noise or an increase in detector contamination.
Reassemble the FID, light the flame, and allow the detector temperature to
equilibrate at 10°C – 50°C higher than the column will reach during typical
operation. This will reduce the amount of phase condensing onto the detector
parts.
Do not exceed the maximum temperature limit of the
stationary phase. Many columns fit far enough into the detector to expose the
phase to these elevated temperatures. Set the proper flow rates for hydrogen
and air (Figure 2), and ignite the flame. Turn on the electrometer and allow a
few minutes for warm-up. The flame should now be stable and noise-free.
Reducing detector noise and contamination
Most detector noise and contamination is the result of
column bleed. The amount of bleed is greatest when the column is initially
conditioned. Your detector will remain clean longer if you condition a new
column before connecting it to the detector. By-products eluted during
conditioning, potentially harmful to the FID, are voided into the oven.
Connect the column inlet to the injector as usual and set
the optimal column flows. Purge the column with carrier gas at an ambient
temperature for an hour before beginning the temperature program. Do not allow
a combustible carrier gas such as hydrogen, and methane to exit the column into
the oven.
Consult the column manufacturer for conditioning details,
i.e., duration and temperature of conditioning. Do not routinely condition new
columns at the maximum temperature limit of the stationary phase – this will
reduce column life. Connecting a well-conditioned column to a clean FID should
produce good sensitivity. If detector stability quickly degrades, evaluate the
quality of the column and carrier gas.
By following these FID maintenance guidelines you will ensure that your test results remain accurate and reproducible.
Edward J.
Zachowski
Edward J.
Zachowski is President and founder of Alpha Omega Technologies, Inc. (AOT) in
Brielle , NJ. He has a Masters Degree in Analytical Chemistry from the State
University of New York at Buffalo, NJ and a Bachelor of Science Degree in
Chemistry from Richard Stockton State College Pomona, NJ. Prior to starting
AOT, Ed was employed by Waters Millipore of Milford MA & Warner
Lambert/Parke Davis R&D Morris Plains, NJ.
Stan
Paleologos
Stan
Paleologos is Laboratory Director at Alpha Omega Technologies, Inc. (AOT) in
Brielle, NJ. He holds a Bachelor of Science Degree in Chemistry from Rutgers
University, New Brunswick, NJ. Prior to joining AOT, Stan served as Laboratory
Manager for various firms including Environmental Management Services Div of
Camin Cargo in Linden, NJ, Blue Marsh Laboratory and Chyun Associates, both in Princeton, NJ.


